NIWOT02: Logbook Entries

NIWOT02: CO2 Messages: 14 Entries..

Return to Logbook Contents Page
Entry Date Title Site Author #Graphics
29 Mon 07-Oct-2002Hydra visit/ teardowndelany
28 Mon 07-Oct-2002Hydra visitdelany
27 Mon 07-Oct-2002Time delays for the Hydra intakesalldelany
26 Mon 07-Oct-2002Final Intercomparison using Brittscalstn2_centraldelany
25 Mon 23-Sep-2002LICOR INTERCOMPARISON SYSTEMstn2_centraldelany
23 Tue 17-Sep-2002Hydra Check liststn2_centraldelany
19 Thu 12-Sep-2002Cal cycle changestn2_centraldelany
18 Thu 12-Sep-2002Hydra visit and Check liststn2_centraldelany
17 Tue 10-Sep-2002Hydra visitstn2_centraldelany
15 Wed 04-Sep-2002Hydra sample period increased againstn2_centraldelany
14 Wed 04-Sep-2002Hydra sample period increasedstn2_centraldelany
13 Wed 04-Sep-2002Hydra operationalstn2_centraldelany
12 Wed 04-Sep-2002fixed opto22 command problemstn2_centralmaclean
10 Tue 03-Sep-2002opto22.s2stn2_centralmaclean


29: CO2, Site , Mon 07-Oct-2002 16:17:21 MDT, Hydra visit/ teardown
30 Sep''02, Arrived 10:05
Steve S, Kirt, Tony, John, Steve O, also Jielun

All three pumps OK
Descicant 2/3 pink, 1/3 blue
All aspiration flouting ball between 6.0 and 6.5

Licor A: sample/cal 6.5 in cal
         dry air 6.5
         Ref 7.5 
         P = 1.2 when pump off when in calibrate mode
           = 5.7 when in sample mode

Licor B:  sample/cal 6.5+ in cal 
          dry air 6.5
          Ref 3.0
          P = 1.4 when pump off in cal mode
            = 5.4 when in sample mode
Licors powered off at 10:21
No Brittcal gases

Ref 1500/5.6 no flow        LK 438863
Cal2 1400/5.9 no flow       270552
Spare ref capped            REE 222
Cal3 1660/5.8 no flow       K361224
Cal1 1510/6.1 no flow       414576

All Hydra components removed.
except the intake lines. The lines into the hydra box were clamped tightly 
in the strain relief. The top lines, A 1-9, were tagged with black panduit ties.




28: CO2, Site , Mon 07-Oct-2002 16:09:49 MDT, Hydra visit
24 Sep''02, Tuesday
Tony Delany and Don Lenschow
Both RAA pumps and dry air pump OK
All aspiration floating balls between 6.0 and 6.5. All adjusted to 6.25
Ref  1380/4.2
Cal2 1520/6.2
Spare ref Capped
Cal3 1750/5.8
Cal1 1630/5.6

Licor A: sample/cal on sample 65
         dry air 62
         ref 5, increased to 6
         P = 5.8 on sample
Licor B:  sample/cal on sample 64
          dry air 60
          ref 2, increased to 6
          P on sample 5.1

Hydra delay times determined. See other LOGBOOK entry
27: CO2, Site all, Mon 07-Oct-2002 10:07:33 MDT, Time delays for the Hydra intakes
24 Sep''02. Tony D and Don L visited Niwot02 site and undertook to measure the
time delays for the Hydra.

The AC power for the relays, and hence solenoids, was switched off. 
The AC power for the Opto22 and the 12 VDC for the sample valve and the 
sample pump were maintained.
The AC daisy-chain and the corresponding connections from the solenoids 
were disconnected. Zip cords were connected in their stead. Thus on the flip 
of a switch on the power block a single solenoid could be activated.

The flow through the buffer volumes was continuous and the choice of 
solenoid merely changed the choice of sampled volume. 
The solenoids were activated at the instant of an Opto22 transition and the 
instruction to breathe into the intake was also given, by radio at the 
instant of an Opto22 transition. Generally the buffer volume was sampled 
for 80 x 2 seconds befor the breath instruction was given. 
then 30 seconds of breathe was sampled. A 10 - 15 minute period was allocated 
to allow the pulse of CO2 to travel the length of the Hydra intake line.
 A table of delay times is gien in 
delany/deployments/niwot02/hydratimes.fm
26: CO2, Site stn2_central, Mon 07-Oct-2002 09:47:22 MDT, Final Intercomparison using Brittscal
9/26/02 INTERCOMPARISON SYSTEM FIELD NOTES
Sean Burns took these data.

       Tank Pressures with Valves
       Cylinder    Closed     Open
       --------    -------    ----- 
       A             1200     1200
       B             1320     1350
       C                0     1250
       D             1600     1600


       System       Time*     Event               IC Gas    Line
       ------      --------   -------------      --------  ------
       NCAR-A      14:24:45    Connect to LiCorA  -- Hydra was on S5 
       NCAR-A      14:25:30    step 4-way           A
       NCAR-A      14:28:30    step 4-way           B
       NCAR-A      14:31:30    step 4-way           C
       NCAR-A      14:34:30    step 4-way           D
       NCAR-A      14:37:30    3-way back to hydra          hydra on S6.

       NCAR-B      14:41:02    3-way to ICS         -       
       NCAR-B      14:41:30    step 4-way           A  
       NCAR-B      14:45:00    step 4-way           B
       NCAR-B      14:48:00    step 4-way           C
       NCAR-B      14:51:00    step 4-way           D
       NCAR-B      14:54:00    3-way back                   hydra on ?

       Time Check:  at 14:38:46 switch from s6-s7
       Note: Hydra Cal was just finishing as i arrived (~14:20).

       -------
       USGS Time Check-- 14:20:00 USGS = 15:20:04 (watch time, which was
                                             about 3 sec behind GPS time)

  USGS        15:17:20   3-way to ICS         on L5, P~509 torr
  USGS        15:18:30   run start            A  P?, CO2~431.1
  USGS        15:21:34   step 4-way           B  P~546 torr, CO2~403.8
  USGS        15:24:34   step 4-way           C  P~518 torr, CO2~389.5
  USGS        15:27:34   step 4-way           D  P~450, CO2~457.0
  USGS        15:28:30   still on D --> increased P to be near 515 torr.
  USGS        15:31:34   3-way back   USGS Time=14:06:01  on L7


  CU          16:09:40   to man, C2 on                 L4
  CU          16:11:24   3-way to ICS         
  CU          16:12:00   step 4-way           A   (P~8 psi)
  CU          16:16:00   step 4-way           B   (P~12 psi)
  CU          16:20:00   step 4-way           C   (P~24 psi)
  CU          16:24:00   step 4-way           D   (P~18 psi)
  CU          16:25:23   still on D,  increased P to 20 psi.   
  CU          16:28:00   step 4-way           A
  CU          16:30:00   step 4-way           B
  CU          16:32:00   step 4-way           C
  CU          16:36:00   3-way back to CU
  CU          16:37:15   C2 on, to auto            L1

  Time Check:  at 16:38:24 switch from L2-L3


        *times shown are from my watch which is off from GPS by about
         3-4 seconds...ie, GPS= 19:41:00, watch=19:41:03.

                     --------------------------------

  All the tank pressures and flows on the Hydra looked fine.  
...all the Britt intercomparison details are also posted on the MMM niwot
ridge webpage under "Project Timeline and Updates".  ie,

          http://www.mmm.ucar.edu/abl/forest/

                      
25: CO2, Site stn2_central, Mon 23-Sep-2002 11:32:02 MDT, LICOR INTERCOMPARISON SYSTEM
9/19/02 INTERCOMPARISON SYSTEM FIELD NOTES
Sean Burns took these data.

        Tank Pressures with Valves
Cylinder    Closed     Open
--------    -------    ----- 
A                0     1250
B             1400     1490
C                0     1300
D                0     1610


System       Time*     Event               IC Gas    Line
------      --------   -------------      --------  ------
CU          15:00:28   to man, C2 on                 L1
CU          15:02:01   3-way to ICS         
CU          15:02:35   step 4-way           A  (P~4 psi)
CU          15:05:00   step 4-way           B  (P~12 psi)
CU          15:07:30   step 4-way           C  (P~20 psi)
CU          15:11:00   step 4-way           D  (P~20 psi)
CU          15:14:30   3-way back to CU
CU          15:16:00   C2 on, to auto                L4

USGS        15:33:30   3-way to ICS  USGS Time=14:33:30  on L10
USGS        15:33:40   run start            A  P~4, CO2~429.7
USGS        15:37:00   step 4-way           B  P~4, CO2~404.4
USGS        15:40:00   step 4-way           C  P~10, CO2~390.1
USGS        15:43:00   step 4-way           D  P~6, CO2~450.9
USGS        15:46:01   3-way back    USGS Time=14:06:01  on L7

NCAR-A      16:22:30   Connect to LiCorA  -- Hydra was on S4 
NCAR-A      16:23:30   step 4-way           A
NCAR-A      16:27:00   step 4-way           B
NCAR-A      16:29:30   step 4-way           C
NCAR-A      16:32:00   step 4-way           D
NCAR-A      16:35:00   3-way back to hydra          hydra on S2 (?).

NCAR-B      16:37:00   3-way to ICS         A       hydra on S6.
NCAR-B      16:38:00   step 4-way           A  (slight pressure adjust)
NCAR-B      16:41:00   step 4-way           B
NCAR-B      16:43:30   step 4-way           C
NCAR-B      16:46:00   step 4-way           D
NCAR-B      16:48:30   3-way back                   hydra on S4

Time Check:  at 17:00:03 switch from s4-s5
             at 17:01:23 switch from s5-s6


*times shown are from my watch which has been adjusted to be within
1 sec of GPS time.

--------------------------------
23: CO2, Site stn2_central, Tue 17-Sep-2002 09:15:27 MDT, Hydra Check list
 
                              HYDRA CHECK LIST

Both RAA Gast pumps operating?
Dryer pump operating?

All aspiration floating balls between 6 and 6.5?. If not then adjust.

Note cylinder pressure and regulator pressure
Ref                 (eg 1750/4.9)
Cal 2               (eg 1560/5.9)
Spare Ref              capped
Cal 3               (eg 1830/6.2)
Cal 1               (eg 1760/6.1)

Hydra box: open top and watch the little lights to see that the cycle is 
progressing properly:
 ... cal3, cal2, cal1, Sample/s9,s8,s7,s6,s5,s4,s3,s2,s1, cal3, cal2, etc ...

Look at flow tubes on inside of door. Note values and,if necessary,adjust knob
     below guage to set at correct value.
Look at pressure at pump output. Note value and, if necessary, adjust the knob
     on the pressure controller T'd with the pressure guage.
Licor 1: Dry air                 =            (should be 65   )
         Ref                     =            (should be  5   )
         sample/cal              =            (should be 66   )
         P (when in sample mode) =            (should be  6.0 )

Licor 2: Dry air                 =            (should be 65   )
         Ref                     =            (should be  5   )
         sample/cal              =            (should be 66   )
         P (when in sample mode) =            (should be  6.0 )
        

Both tubes of desiccant.  Blue OK, If totally pink, replace.

Undertake a Brittcal on both Licors. 
Keep track of the "Sample" / "Calibration" sequence.
Fit each entire Brittcal into a single "Sample" period as overlapping with a 
"Calibration" causes some problems.
    

19: CO2, Site stn2_central, Thu 12-Sep-2002 16:17:51 MDT, Cal cycle change
Gordon changed the calibration period and duration and the reset interval
At 14:19:
       the calibration period was changed from every 30 min to every 60 min.
       the calibration durations were changed from 2 min to 4 min                      the reset interval was changed from every 2.5 hrs to 5 hrs.
These changes do not modify the total cycle time of 1 hour.

At 16:15 this was checked back at FLAB and seen to be so.

18: CO2, Site stn2_central, Thu 12-Sep-2002 11:44:39 MDT, Hydra visit and Check list
Steve O and Tony D arrived at the NCAR tower at 11:30
RAA aspiration, all balls between 6 and 6.5
Cal2 valved off 1560/ 19 psi
Ref             1740/5 psi

The new AC cal valves had been plumbed in the day before but not electrically.
At 11:54 AC to all black relays off. (AC to Opto22 left on)
Electrical reconnection (pink relays replaced by black, DC jumpers removed, 
AC jumpers installed, TM10 wiring to Licor box barrier strips disconnected, 
New valve wiring connected to barrier strips) completed and AC power restored 
at 12:45.

12:24 Licor 1 P = 5.8 sample/cal = 65
      Licor 2 P = 5.0 sample/cal = 65

13:15 Licor 1 Britt-cal CalA regulator adjusted to bring Britts intake P to 5.8
                        CalB                  ditto
                        CalC                  ditto
                        CalD                  ditto
13:35 Licor 2 Britt-cal CalA regulator adjusted to bring Britts intake P to 5.0
                        CalB                  ditto
                        CalC                  ditto
                        CalD                  ditto
    Note the Britt-cal for Licor 2 interupted by Cal cycle. Extended Britt-cal

14:02 Filled descicator tubes with blue Dryrite, powered pump and adjusted 
     dry air flows to 6 for both Licors

Finished removing all the ss nipples on front of intake filters.

14:32 Returned from last of filter nipple removals
                              CHECK LIST

Both RAA Gast pumps OK
Dryer pump OK

All aspiration floating balls between 6 and 6.5

Ref                 1750/4.9
Cal 2               1560/5.9
Spare Ref           capped
Cal 3               1830/6.2
Cal 1               1760/6.1

Hydra box: open top and watch the little lights to see that the cycle is 
progressing properly:
           Sample/s1,s2,s3,s4,s5,s6,s7,s8,s9  , cal1 , cal2 , cal3 , 

Licor 1: Dry air =                 55
         Ref =                      5 ( incr to 7)
         sample/cal =              62 ( incr to 65)
         P (when in sample mode) =  5.8

Licor 2: Dry air =                 58
         Ref =                      5 
         sample/cal =              68
         P (when in sample mode) =  5.0

Descicant  Blue OK OK

14:46    Leave
    










17: CO2, Site stn2_central, Tue 10-Sep-2002 10:57:23 MDT, Hydra visit
Tony D and Britt S visited Niwot site on Thurs 5 Sep''02
Arrived 14:50 

Both Gast RAA Hydra aspirator pumps running.
All floating balls  between 550 and 650. Adjusted all balls to 60 +_ 3
on the 102-05N (st). This corresponds to 3100 ccpm uncorrected for P and T.

Gas cylinders, Ref  1850/4 psi
               Cal2 1700/7.2 psi

15:00 Inlets swopped A1 <-> A4. This was done to bring the 
Licor A NCAR 1m sample into time synch with the Licor B NCAR 1m sample.

Licor A sample/cal flow = 50 on N032-15(st), increased to 65 ( =  145     ccpm)
Licor B sample/cal flow.= 40 on N032-15(st), increased to 65 ( =  145     ccpm)
NB caltable for N032-15 (ST):
05 =8, 10 =12, 20 =27, 30 =45, 40 =65, 50 =97, 60 =125, 65 =145 ml min-1

Licor A ref flow = 3 on N032-15(st), increased to 5 ( =    8   ccpm)
Licor B ref flow.= 4 on N032-15(st), increased to 5 ( =    8  ccpm)

Licor A KNF pump P = 5.5 psi
Licor B KNF pump P = 4.5 psi

During the period 15:33 to 15:38 observed sequence sampl/cal1/cal2/cal3/sample.
Looked OK

!6:40 - 16:51 Britt's calibrator to Licor B, ran CalA, CalB, CalC and CalD
!6:53 - 17:03 Britt's calibrator to Licor A, ran CalA, CalB, CalC and CalD



15: CO2, Site stn2_central, Wed 04-Sep-2002 13:08:41 MDT, Hydra sample period increased again
At 13:08 MDT the Hydra sample period was increase from 40s to 80s
The number of cycles between cals was decreased from 4 to 2 thus maintaining 
the total cal-cal as constant.

14: CO2, Site stn2_central, Wed 04-Sep-2002 11:09:46 MDT, Hydra sample period increased
At 10:41 the Hydra sample period was increase from 20s to 40s
The number of cycles between cals was decreased from 8 to 4 thus maintaining 
the total cal-cal as constant.
13: CO2, Site stn2_central, Wed 04-Sep-2002 10:45:43 MDT, Hydra operational
On Thursday, 28 Aug''02 the Licors were installed with the Hydra and the system set to run over the weekend.The data inspected on Tuesday showed little 
variation. The system apparently was not cycling through the different intakes.
On 3 Sept, Tuesday the system was inspected. 
Problems: 
-the DC had been disconnected when the DC power was run through the BOB. 
Thus theOpto22 relay for the sample valve and the sample pump operated but 
here was  resulting operation. Power restored.
-the serial message from ther ADAM was not received properly and so the opto22 
ran intermittently sometimes operating for several cycles and the switching off and restarting. Gordon changed the buffer timing and the system operated well.

Inspection of the data on Wednesday indicated that the Hydra operated properly 
overnight.
Promblems:
- the calibration is still incomplete owing to the single cal gas feeding the 
manifold valving. This will be rectified when the new manifold valves are 
installed.
- there is apparently a 7-8 sec delay after sample valve change before the 
analyser responds.
-there is not time within the 20 sec period for the analyser to come to steady
state with the new sample.
The latter two issues will be addressed by increasing sample flowrate, 
shortening the sample manifold to analyser lines and increasing the sampler
period.


12: CO2, Site stn2_central, Wed 04-Sep-2002 09:12:10 MDT, fixed opto22 command problem
When Tony was up at the hydra yesterday he saw the opto22 relays behaving
erratically.  It would work for 5 minutes or so, then quit, then work again.

I think the problem was in rserial.  

If rserial's input is not from a terminal (in this case rserial's 
input is from a pipe), rserial was adding its own 2 second delay before
sending the data. rserial is also not reading in chunks of lines,
it is just reading what is available.  Because of these two
factors I believe a series of opto22 commands like this:

	>00J2001??
	   20 seconds
 	>00J2002??
	   20 seconds
	>00J2004??

could be received at the opto22 like this:
	>00J2
	   2 seconds...
	>001??
	   18 seconds
	>00
	   2 seconds ...
	J20
	   2 seconds...
	02
	   16 seconds

I think this confused the opto22.

The delay has been removed from rserial, and from the CO2 data, it
looks like the opto22 has been working correctly since then:
Sep 3 22:37 GMT.




10: CO2, Site stn2_central, Tue 03-Sep-2002 14:18:38 MDT, opto22.s2
The variable opto22.s2 contains the value of the last 4 hexadecimal
digits of the command ">00JXXXX", that was sent to the opto22 valve 
controller  on the hydra.

Note that opto22.s2 is value of the command that was sent. An
indication of whether the opto22 actually received the command
and activated the given relays is not directly available in the
present configuration.


To see which relays were switched convert opto22.s2 to hex. The bits
indicate which relays.

  opto22
decimal hex     meaning
8193    2001    main sample valve (2000hex) and first valve open
8194    2002    sample and second valve open
8196    2004    sample and 3rd valve
8200    2008    sample and 4th
8208    2010    sample and 5th
8224    2020    sample and 6th
8256    2040    sample and 7th
8320    2080    sample and 8th
8448    2100    sample and 9th valve open

4608    1200    main cal valve (1000hex) and 1st cal valve open
5120    1400    main cal and 2nd cal valve open
6144    1800    main cal and 3rd cal valve

0       0       opto reset (every 2 hours), command=">00B"
-65535  FFFF    opto config, command=">00GFFFF",
                all ports configured for output