Daily Status, Wednesday, June 19 Looking at past 2 days, ending at 0900 CDT. Staffing: Militzer & Semmer Recent maintenance: s7: visit yesterday to download SDR data s2: install SDR system at 2 s1: fix GOES problem late yesterday Anomolies: s5,6,9 now have single-level Qsoil profile outages s6 krypton noisy (GOES?) s7 krypton still low variance s6&s9 should have UW sonics s5 : down at ~3:00CDT, came back to life around 7:00CDT Wx -- Spd ok Dir ok T ok RH ok P ok Rain questionable rain at 1, 4, and 6 Radiation --------- Rnet ok Rpar ok Rsw(.in,.out) ok Rlw ok Rpile ok albedo - skipping Tdome-Tcase - ok Tdome-T - ok Tcase-T - ok Tsfc,derived=F - ok Tsfc vs Tsfc - skipping Rnet vs Rsum - ok, daytime diff 20 W/m^2 Sonic ----- spd - ok dir - ok w - ok tc - ok Tc vs tc - skipping u'u' - ok v'v' - ok w'w' - ok tc'tc' - ok u* - skipping sigma_w/u* - skipping Cd ? w'tc' - ok sigma_tc/tc* - skipping Scalars ------- h2oV s5 very low signal, took a drop at yesterday morning h2o ok h2o vs H2O - skipping h2o'h2o' s6 noisy (GOES?); s7 low [now collecting fast data to permit a correction in post-processing] w'h2o' see above; s7 low co2mr ok co2mr'co2mr' ok w'co2mr' ok Near-surface Soil ----------------- Tsoil ok Qsoil ok, s7 and s9 very high compared to others Gsoil ok, s1 bad Soil Profiles ------------- par(mfrow=c(3,5)); for (i in 1:13) plot(dat("Tsp",profile=i)) Tsoil ok, p12(s5),lev5 noise yesterday afternoon Qsoil problems: p1(s6),lev4 spikes; p8(s9),lev3; p9(s9),lev1; p12(s5),lev6 noisy; p13(s9),lev1,5 WPtemp ok WPdiff p9(s9),lev1; p10(s7),lev1, p13(s9),lev2 ? Power ----- (2 & 7 have old style power monitors) Vbatt ok Tbatt ok Iload ok Icharge ok GOES ---- Signal ok ClockError ok FreqOffset all within 100 Modulation - skipping Data Quality - skipping
The net radiometer Q99258, which had been removed from site 2 on 13 Jun''02, was repaired and replaced in the spare sensor pool. The upper dome had been dimpled by heavy rain or hail. It was replaced with a new dome from the cache of net radiometer domes in the base.
Arrived at site 10 1/2mile west and 2 miles north of site #3, at 18:40 Obtained some dubious clods. Leveled the radiometer Departed at 19:10
Serializer and SlowAD were checked against a 'dial-a-volt' reference for consistency. Reference: 2.3995 (based on 4.5 digit dvm. output varied slightly between boards from about 2.3993 to 2.3998) A/D Values reported are nominal of what was observed. A/D Consistency: The 4 channels on each pmb-ad board were consistent within a few counts between the channels and between readings. Serializer Values: The measurement was that reported by EVE after scaling of the raw readings. These include the divide-by-2 cable used to scale the krypton voltage and represent mV. Note that for every unit, the nominal values shown had noise of about "1" mV total during the brief test. Serial Numbers: Dash numbers represent the residual reference voltage reading taken by Steve Oncley used in adjusting the output. For example"-07" represents 2.5007, "-93" represents 2.4993. PMB boards were tagged "A1" and "S1", etc. Sta s/nAD s/nSerial AD0 AD1 Serializer 1 A1-93 S1-07 23988 23987 2385 Note 1) 2 A2-95 S2-02 23976 23962 2390 3 A3- S3- 23987 23988 2390 4 A4-04 S4-96 23988 23987 2392 5 A5-88 S5-94 23985 23983 2391 6 A6-91 S6-09 23985 23981 2392 7 A7-06 S7-88 23984 23980 2392 8 A8-94 S8-99 23976 23976 2395 9 A9-97 S9-86 23985 23987 2391 Note 1) Station 1 was sitting in the cal-lab adjacent to the running co2 system. There was quite a bit of noise on the a/d counts while taking the readings inside to box there. The readings were roughly 23740/23770 and were changing by up to 100. The readings shown above were taken in a different box in a different location and the a/d counts were extremely consistent. I did not repeat the testing to determine whether the noise was endemic to station 1's box or not. Note that the readings for station 8 were taken in the same location.