Project #2002-301 IHOP_2002

International H2O Project 2002

D. Parsons, T. Weckwerth, et al.

U.S. Naval Research Laboratory P-3 Orion


Summary of Calibration Factors


 

Pre-Calibration - 05/01/2002
Variable (Units) Variable Name C1 C2 C3
Total Temperature-Radome Right (C)TTRR-9.6305.046-0.0090
Dew Point Temperature-Top (C)DPT0.25010.1160.0000
Vertical Press. Diff.-Radome (mbar)ADIFR0.3405.211-0.0007
Horizontal Press. Diff.-Radome (mbar)BDIFR0.3305.325-0.0022
Dynamic Pressure-Radome (mbar)QCR69.3806.901-0.0012
Dynamic Pressure-Fuselage (mbar)QCF86.3808.6190.0001
Static Pressure-Fuselage (mbar)PSF541.18054.216-0.0110

 

Post-Calibration - 08/05/2002
Variable (Units) Variable Name C1 C2 C3
Total Temperature-Radome Right (C)TTRR-9.6005.056-0.0093
Dew Point Temperature-Top (C)DPT+0.1609.8680.0012
Vertical Press. Diff.-Radome (mbar)ADIFR0.7005.203-0.0003
Horizontal Press. Diff.-Radome (mbar)BDIFR0.4415.279-0.0025
Dynamic Pressure-Radome (mbar)QCR69.3906.897-0.0010
Dynamic Pressure-Fuselage (mbar)QCF86.2738.6270.0005
Static Pressure-Fuselage (mbar)PSF540.84954.222-0.0076

Differences between the Pre- and Post-Calibrations were within acceptable levels, thus indicating generally stable response. Pre-Project calibrations have been used in the final data processing.

+ Certain sensors had to be replaced during the course of the experiment. Post-Calibration coefficients were used from the point of the exchange to the end of the project.

 

Calibration Maneuvers 05/27/2002, 06/15/2002, 06/18/2002

RAF uses sets of in-flight maneuvers specifically designed to help evaluate the performance of the 3-D wind measurement system on the aircraft. An additional set of maneuvers was flown during test flight TF03 (05/18/2002). Measurement errors for the wind variables were within normal performance standards.


RAF's IHOP_2002 Page | RAF Home Page | EOL Home Page | NCAR Home Page