Graphic: |
Look at Graphics File for serial numbers at Site 9
13: Radiation, Site all, Wed 30-Jul-2003 09:39:06 MDT, Site `8 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 8
- 12: Radiation, Site all, Wed 30-Jul-2003 09:38:41 MDT, Site 7 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 7
- 11: Radiation, Site all, Wed 30-Jul-2003 09:38:09 MDT, Site 6 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 6
- 10: Radiation, Site all, Wed 30-Jul-2003 09:37:36 MDT, Site 5 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 5
- 9: Radiation, Site all, Wed 30-Jul-2003 09:36:14 MDT, Site 4 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 4
- 8: Radiation, Site all, Wed 30-Jul-2003 09:35:23 MDT, Site 3 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 3
- 7: Radiation, Site all, Wed 30-Jul-2003 09:34:00 MDT, Site 2 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 2
- 6: Radiation, Site all, Wed 30-Jul-2003 08:57:30 MDT, Site 1 Rad/Soil Configs / ID's
- Previous -
Next -
Index
Graphic: |
Look at Graphics File for serial numbers at Site 1
- 3: Radiation, Site all, Tue 24-Jun-2003 14:01:51 MDT,
After all 18 pyrgeometers had been exposed to the NOAA Black Body
they were returned to the cal-lab on Fri 20 Jun.
On Tues 24 Jun Lou' four were returned to his lab and all 14 of our's
were opened and the jumpers reinserted.
NB 29137 had been modified such that the Case temp is now defined by the
compensation circuit's case thermister.
- 2: Radiation, Site all, Wed 11-Jun-2003 14:43:11 MDT,
Longwave, pyrgeometers prepared for Black Body pile calibration:
jumpers disconnected.
Schematics of jumper configuration at delany/deployments/hvams/pyrgcalprep.idr
8 Al case ISSF standard config
31974 - Tue
31975 ------- Wed
31976 - Tue
31977 ------- Wed
31978 ------- Wed
31979 ------------ Thu
31980 ------- Wed
31981 - Tue
2 Bronze case ISSF-standard config
26416 - Tue
29260 ------- Wed
4 Bronze case non-ISSF-standard config
26417 - Tue
27907 - Tue
29136 ------- Wed
29137 ------- ----- Thu
In addition 4 Bronze case Eppley-standard pyrgeometers of Lou Verstraete
will be calibrated. These require no jumper modification
23576 ------------- Thu
28809 ------------- Thu
32686 ------------- Thu
32687 ------------- Thu
All the desiccant cannisters were refilled.
Each day the standard plus the six pirgs of the day were exposed to the
black body cavity in the indicated order over the temp range ~ -40 to + 15
utilizing the proceedure of Ellesworth Dutton.Approximately 5 cycles were
performed throughout the 9 hours from 08:00 to 17:00
Tony did Tues, Cody did Wed, and Raychel did Thu
#29137 gave a bad case temperature. The pyrg was brought back to the lab.
The case thermister was checked and found to be bad. The compensation case
thermister was wired in as a substitute
Tue Wed Thu
17 Jun 03 18 Jun 03 19 Jun 03
24333 Std 24333 Std 24333 Std
31981 31975 29137
26417 31980 31979
31974 29136 28809
26416 31978 28810
31976 29260 32686
27907 31977 32687
Repeat Repeat Repeat
|
|
|
|
|
|
|
|
|