Haven't run system for a while, so start N2 flow without starting a loop. Bypass flow @ 9:09 Time Meeco Reeading Meeco In(V) .vbp H2O (ppm) Notes 9:09 26.7 4.51123-4.51786 26.25-26.336 Bypass flow on 30.0 4.53565 27.0 24.5 4.48179 24.350 23.5 4.46531 23.34 23.0 4.45843 22.9 19.0 4.39019 17.89 17.5 4.36479 17.19 9:13 10.0 4.24248 9.70 Bypass and sample flow 8.0 4.21322 7.91 7.0 4.19720 7.05 6.0 4.18107 5.976 5.0 4.16418 5.105 9:18 4.0 4.15118 4.113 9:21 3.0 4.13553 3.155 9:29 2.0 4. 4.11789-4.11932 2.087-2.312 10:00 1.3 Flow off N2 loop starting @ 10:31 Connect ND19517, vent x 4. Pressurize, let sit ~ 1 hr, vent again. Close N2 ball valve, open Bypass flow until N2 @ 0 psi, close bpass flow valve, switch 4 way and open ND19517. Autoloop @ 12:34. N2 autoloop N2 autoloop @ 15:13 Watched as N2 loop goes into No Flow mode: Meeco reading goes quickly from 1.0 to 2.7ppm