AO2: opened lamp box in preflight - coil pin looks fine - will pay attention to pd noise to see if that is the cause of the motion noise CO2 jumped by ~ -10 ppm on take-off (related to tray in/out??) went to calint = 1, ltfreq = 1, wtfreq = 1 to rerun 4-pt cal 20:21 forgot to undo, thought it had started another cal - went to MAN, SA sel, SA prg off, but it was still doing first loop - ended up cutting WT short Set WTint to 1 so it will do WT next time 20:41 starting down from 40 kft, went back to AUTO and calint 1 to force cal 20:52 wtfreq back to 4 21:01:00 bumps, o2n = 5, pdo2n = 0.6, pdspn = 5, pdwtn = 11 21:02:30 bumps, o2n = 6, pdo2n = 1.0, pdspn = 7, pdwtn = 11 21:03:20 bumps, o2n = 7, pdo2n = 1.0, pdspn = 7, pdwtn = 14 21:12:00 smooth, o2n = 3, pdo2n = 0.5, pdspn = 4, pdwtn = 6 21:13:30 smooth, o2n = 2, pdo2n = 0.4, pdspn = 5, pdwtn = 6 21:21:00 bumps, o2n = 5, pdo2n = 0.4, pdspn = 6, pdwtn = 14 21:13:30 smooth, o2n = 4, pdo2n = 0.3, pdspn = 5, pdwtn = 7 22:30 extreme drift on LT 22:42 and 22:43 CO2 jumped twice by a total of ~ 6 ppm 23:17:52 to MAN, SA sel, SA purge off, b/w HS and LS to get low point 23:25:53 back to AUTO, starts w/ HS purge 23:25 another drifting LT 00:46:00 smooth, o2n = 1.7, pdo2n = 0.2, pdspn = 6, pdwtn = 7 (IN REAL AIR) 01:32:00 smooth, o2n = 1.9, pdo2n = 0.26, pdspn = 5, pdwtn = 7 (IN REAL AIR) 02:44 set calint to 500 to delay cal until top of profile ??:??:00 bumps, o2n = 4, pdo2n = 0.5, pdspn = 5, pdwtn = 10 03:22 calint to 1 to force cal then 50 03:22:00 bumps, o2n = 4, pdo2n = 0.3, pdspn = 5, pdwtn = 8 between 2 and 3:30, CO2 cal went up by 1 ppm - SA doesn't show a jump but maybe a slow trend 04:14:20 to MAN, unpurge LS 04:15:17 to WT CO2 jumped by 90 ppm on touch down MED: new code was not purging 30-32, went back to old code flushing flask 1, P1, PSA, and Flow dipped 3 times - upstream pressure (having trouble controlling at 40 kft) 03:34 - have averaged a little over 1 % water the entire flight - done sampling, but will let pumps run on bypass until landing to try to plug traps 1% x 8 hours x 3 SLPM = 1440 L / 22.4 L/mole = 64 moles air * 1% = 0.64 moles H2O * 18 g/mole = 11.5 g / 1 g/cc = 11.5 cc H2O actually > 1%, > 8 hours (8:10 eta), < 3 SLPM (now at 1.7) 3:51:43 back to 135/370 good flow to end