AO2: O2 very noisy - WT jogs are uneven and O2d spikes low: - suspect ice blockage at WT U-tube b/c it looks similar to noise on TF02 when ice was in SP U-tube, except spikes are in opposite direction - had WT line open to replace cylinder 2 days ago, could have introduced H2O at that point - will dry out during tomorrow's maintenance day [turned out being a flow problem - see notebook] 00:10 system started purging LS but never switched to LS, watchdog failed ~ 00:18, noticed ~ 00:32, to MAN mode, manually switched a valve and flows restarted, manually relit lamp, to AUTO ~ 00:35 and system appears to cycle normally - somehow, pulling vac for 10 minutes appears to have fixed the O2 noise problem - now around 5 per meg -03:20 o2 return of downward spikes. 04:09 O2 noise back in full force, disabled ch'over with SP to cell and closed WT 248 to pump out for ~2 min. Opened WT 248 and renabled ch'over. Did not seem to help. 04:14-04:14:30 toggled WT 248 ~ 10 times to try to blast out any U-tube blockage. Renabled ch'over. Did not seem to help. Disabled ch'over with WT to cell - no spikes, confirming noise is pressure/flow/switching related ~04:20-04:23 toggled SP 248 open/closed ~ 7 times. Renabled ch'over..... did not seem to help.... MED: Um... nothing to report...