- 94: SABL, Site vtmx, Thu 26-Oct-2000 17:12:00 GMT, sabl data computer fixed.
Ran fsck and got the SABL data computer going again.
- 82: SABL, Site vtmx, Wed 25-Oct-2000 17:26:07 GMT, Leak fixed on SABL
So far changing the couplings has fixed the leak of
coolant for SABL. Pulse seems really weak. Room got
pretty warm last night.
- 78: SABL, Site vtmx, Tue 24-Oct-2000 17:42:23 GMT, SABL weirdness
Scope exhibiting weird things in the scope, but they seem
to have cleared up. The pulse in the scope is down from yesterday.
Will continue to monitor its with the scope. Data display looks good.
Replaced the couplings in the back of the pump to try and slow down
the leak of coolant. Wish us luck.
- 48: SABL, Site vtmx, Tue 17-Oct-2000 17:13:50 GMT, Changing tape and cleaning window, etc.
Changing tape, making copy, checking drip pan and cleaning window.
- 41: SABL, Site vtmx, Tue 17-Oct-2000 10:10:15 GMT, "Lazer" Weak
SABL lazer got really weak. It appears to be a temperature
problem. I added coolant (1/2 glyc, 1/2 h20) and cranked the AC and
that fixed things.
- 16: SABL, Site vtmx, Sat 07-Oct-2000 21:40:13 GMT, sabl tape and flash bulb changed.
Changed the tape on SABL
Changed the flash bulb because the pulse was getting
very weak. Took 3 hours.
Cleaned window.
Restarted SABL and the display didn't start, had to
shut SABL program down and restart it. That fixed everything.