LI-7500 Data Sheet
|
LI-7500 Specifications
Type: Absolute, open-path, non-dispersive infrared gas analyzer.
Detector: Thermo-electrically cooled lead selenide.
Bandwidth: 5, 10, or 20 Hz, software selectable.
Path Length: 12.5 cm (4.72").
Operating Temperature Range: -25 to 50 °C (-40 to 50 °C verified optional range is also available).
User Interface: Windows® based software supports all setup, configuration and calibration functions through the RS-232 serial port.
Outputs: RS-232 (20 Hz Maximum)
SDM (>40Hz)
2 user scalable 16 bit DACs updated at 300 Hz.
Auxiliary Inputs: 2 channels for temperature and pressure sensors (during calibration).
Auxiliary Input with pressure sensor: 0-4.096V (±5V common mode rejection).
Power Requirements: 10.5 to 30 volts DC.
Power Consumption: 30 W during warm-up, 10 W in steady state.
Dimensions:
Head: Dia 6.5 cm, Length 30 cm. Designed for minimal wind flow disturbance.
Control Box: 35 cm x 30 cm x 15 cm (external dimensions).
IRGA cable: 3 meters (between sensor head and electronics control box).
Power, Serial, DAC, Auxiliary Input and SDM cables: 4 meters
Weight:
Head: 0.75 kg (1.65 lb.)
Control Box and Cables: 4.8 kg (10.5 lb.)
CO2 Color
|
----- |
µmol mol-1 |
mmol m-3* |
mg m-3* |
Calibration Range |
----- |
0 - 3000 |
0 - 117 |
0 - 5148 |
RMS noise** at ambient (370 ppm) |
|
0.08 |
0.0031 |
0.13 |
Zero drift with temperature (per °C) |
Maximum
Typical |
±0.3
±0.1 |
±0.012
±0.004 |
±0.5
±0.2 |
Gain drift with temperature at 370 ppm
(% of reading per °C) |
Maximum
Typical |
±0.1%
±0.02% |
±0.1%
±0.02% |
±0.1%
±0.02% |
Direct sensitivity to H2O (mol CO2/mol H2O) |
Maximum
Typical |
±4.00E-05
±2.00E-05 |
±4.00E-05
±2.00E-05 |
±4.00E-05
±2.00E-05 |
H2O
|
----- |
mmol mol-1 |
mmol m-3* |
g m-3* |
Calibration Range |
----- |
0 - 60 |
0 - 2340 |
0 - 42 |
RMS noise** in moist air (20 mmol mol-1) |
|
0.0034 |
0.13 |
0.0024 |
Zero drift with temperature (per °C) |
Maximum
Typical |
±0.05
±0.03 |
±2
±1 |
±0.04
±0.02 |
Gain drift with temperature at 20 mmol mol-1
(% of reading per °C) |
Maximum
Typical |
±0.3%
±0.15% |
±0.3%
±0.15% |
±0.3%
±0.15% |
Direct sensitivity to CO2 (mol H2O/mol CO2) |
Maximum
Typical
|
±0.05
±0.02 |
±0.05
±0.02 |
±0.05
±0.02 |
* at 25 °C, 98 kPa
** 5 Hz bandwidth, 10 Hz sampling |